Calcified plaques demonstrated in this one beat scan reconstructed with PIQE 1024.
View Scan ParametersScan Mode | Volume |
Collimation | 0.5 mm |
kVp | 120 |
mAs | SUREExposure |
Rotation Time | 0.24s |
Scan Range | 130mm |
Dose Reduction | PIQE 1024 Cardiac |
CTDIvol | 13.4mGy |
DLP | 174.4mGy.cm |
Effective Dose | 2.4mSv |
k-factor | 0.014 (AAPM Report 96) |
Aneurysms are seen in the RCA and LAD
View Scan ParametersScan Mode | Volume |
Collimation | 0.5 mm |
kVp | 100 |
mAs | SUREExposure |
Rotation Time | 0.24s |
Scan Range | 120mm |
Dose Reduction | PIQE 1024 Cardiac |
CTDIvol | 8.4mGy |
DLP | 101mGy.cm |
Effective Dose | 1.4mSv |
k-factor | 0.014 (AAPM Report 96) |
PIQE provides clearer delineation of the aortic valve leaflets compared to hybrid IR reconstruction.
View Scan ParametersScan Mode | Volume |
Collimation | 0.5 mm |
kVp | 100 |
mAs | SUREExposure 3D |
Rotation Time | 0.275s |
Scan Range | 140mm |
Dose Reduction | PIQE 1024 |
CTDIvol | 22mGy |
DLP | 351.7mGy.cm |
Effective Dose | 4.9mSv |
k-factor | 0.014 (AAPM Report 96) |
© Canon Medical Systems Asia Pte. Ltd.
© Canon Medical Systems Asia Pte. Ltd.