A non contrast brain scan reconstructed with AICE demonstrates excellent gray white differentiation and noise free images.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 120 |
mAs | SUREExposure |
Rotation Time | 1.0 s |
Scan Range | 185 mm |
Dose Reduction | AiCE |
CTDIvol | 57.3mGy |
DLP | 1322.6mGy·cm |
Effective Dose | 2.78 mSv |
k-factor | 0.0021 (AAPM Report 96) |
A non contrast scan of the neck was reconstructed with AiCE and SEMAR. AiCE dramatically reduces the noise while SEMAR removes the artifact from the dental fillings
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 120 |
mAs | SUREExposure |
Rotation Time | 0.5 s |
Scan Range | 213.0 mm |
Dose Reduction | AiCE |
CTDIvol | 14.6mGy |
DLP | 375.5mGy·cm |
Effective Dose | 1.46 mSv |
k-factor | 0.0039 (AAPM Report 96) |
A non contrast scan of the chest was reconstructed with AiCE. AiCE Lung dramatically reduces the image noise and maintains the spatial resolution of the lung parenchyma
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 120 |
mAs | SUREExposure |
Rotation Time | 0.5 s |
Scan Range | 700 mm |
Dose Reduction | AiCE |
CTDIvol | 4.9mGy |
DLP | 367.7mGy·cm |
Effective Dose | 5.15 mSv |
k-factor | 0.014 (AAPM Report 96) |
A retrospective ECG gated scan of the heart was performed in this patient with a heart rate of 70bpm and reconstructed with AiCE.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 120 |
mAs | SUREExposure |
Rotation Time | 0.4 s |
Scan Range | 135mm |
Dose Reduction | AiCE |
A scan of the abdomen & pelvis was performed in Arterial and portal phases and reconstructed with AiCE.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 100 |
mAs | SUREExposure |
Rotation Time | 0.5 s |
Scan Range | 630 mm |
Dose Reduction | AiCE |
CTDIvol | 8.2mGy |
DLP | 557.1mGy·cm per phase |
Effective Dose | 16.71 mSv |
k-factor | 0.015 (AAPM Report 96) |
A scan of the abdomen & pelvis was performed and reconstructed with AiCE. Arterial phase iodine maps created with Subtraction SCT demonstrate an enhancing tumor in the abdomen. A large collection with inflammatory tissue is seen in the right groin. Both masses are clearly visualized in grey-scale and color SCT.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 80 |
mAs | SUREExposure |
Rotation Time | 0.5 s |
Scan Range | 490 mm |
Dose Reduction | AiCE |
CTDIvol | 3.7mGy |
DLP | 197.4mGy·cm per phase |
Effective Dose | 8.89 mSv |
k-factor | 0.015 (AAPM Report 96) |
A scan of the chest was performed to rule out coronary artery disease and pulmonary emboli. A vHP scan was performed utilizing a retrospective ECG gated scan of the heart and a non gated scan of the remaining chest. Perfusion defects are seen in the iodine maps corresponding to pulmonary emboli.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 120 |
mAs | SUREExposure |
Rotation Time | 0.4 s |
Scan Range | 261mm |
Dose Reduction | AiCE |
AiCE provides excellent low noise images. The bone reconstruction provides excellent spatial resolution.
Courtesy Radiotherapeutic Institute Friesland, the Netherlands
Scan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 120 |
mAs | SUREExposure |
Rotation Time | 0.5 s |
Scan Range | 320 mm |
Dose Reduction | AiCE |
CTDIvol | 10.1 mGy |
DLP | 369.1 mGy·cm |
Effective Dose | 5.54 mSv |
k-factor | 0.015 (AAPM Report 96) |
Fractures of the metatarsals are seen in this scan of the foot reconstructed with AiCE. AiCE Bone removes the image noise while maintaining the excellent spatial resolution.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 120 |
mAs | SUREExposure |
Rotation Time | 0.5 s |
Scan Range | 205mm |
Dose Reduction | AiCE |
CTDIvol | 1.6mGy |
DLP | 40.1mGy·cm |
© Canon Medical Systems Asia Pte. Ltd.
© Canon Medical Systems Asia Pte. Ltd.